Sterm Sterilizer Introduction
Steam sterilizer with large size capacity while offering automatic horizontal sliding door, intelligent control, reliable operation and easy maintenance, which is suitable for the high level customer with large scales. It is developed according to the latest requirements of medical institution and CSSD.
Steam sterilizer can be widely used in biological engineering, health care, animal laboratory, pharmaceutical industry for the sterilization of biological products, containers, sterile clothing, medical equipment, medical dressings and other items
This QSS scries of pulsation vacuum sterilizer utilizes saturated steam as its sterilization medium which releases a large amount of heat in the stage of condensation, and ensure the sterilizing environment in high temperature and at a certain humidity. Structure and Characteristic of the Equipment
Body of the QSS strilizer.
It is designcd and manufactured in accordance with GB150 *the Pressure Vessel. GB8599 The Technieal Requirements of Large Steam Serilier-Automatic Control Protocol". and TSG Rt004 the -Safety and Technology Specifieation for Fixcd Monitor VesseNi and related insetion procedures.
The chamber in horizontat rectangular shape with doublelayer structure is made from S3008
Doot sealing: Pneumatic scaling: both doors are interlocked and completely conform to the requirementsof the
specification
With standard GMP authentication interface.
Strlizer piping sysem: the pipeline is designed with the best configuration. All the spares are from qualified brands at homc and abroad.
Sterilizer control system: SIEMENS PLC and SIEMENS touch screen, this system is easy to implement thc program selection, parameter seting equipment opeation, report processing and other functions. It has prfect safety measures The FO valuc and thc temperature time gives doublc guarantee for selizaion A single parameter control is also available. It hasa prfect sterilization records.
Main Parameters and Utilities
Design Pressure
0.245Mpa
Working Pressure
0.225Mpa
Design Temperature
139℃
Working Temperature
105-134℃
Vacuum
-0.09Mpa
Temperature Equilibrium
≤±1℃
Water Source Pressure
0.15-0.3Mpa
Compressed Air Pressure
0.3-0.7Mpa
Steam Pressure
0.3-0.7Mpa
Power
AC380V, 50HZ
The 3-phase and 5-wire system AC380V,50HZ